How old is the cuff? If it has written on it "single use" and you've been using it a while you may have microleaks which means you need a new cuff. Also, check the hose for holes (usually found where the hose splits and connects to the cuff). Last, check the Dinamap to hose connector are the black orings still on the connecting posts?
Critikon/Ge utilizes the oscillitory method for determining NIBP. In other words, it pressurizes the cuff, stabilizes, decreases pressure slightly, stabilizes, etc. so a leak in the cuff or hose won't allow the pressure to stabilize resulting in an E83 error.